Abstract
Background: 99mTc is a radioactive isotope that is obtained by eluting a 99Mo/99mTc
generator. (PINSTECH, Islamabad) and used for radionuclide scanning.
Objectives: The objective of this work is to study the uncertainties in 99mTc activity that exist due to time
delay between injection preparation and administration to patients, during the process of gamma camera
scanning.
Methods: Lead canisters were used for storing elution vials and dose calibrator for measuring 99mTc
activity in mCi. The activity of preparing 99mTc injection and its administration to patients were compared
with the prescribed values of activity recommended in the Society of Nuclear Medicine procedure
guidelines.
Results: This study showed that uncertainty in the activity existed in one thyroid patient, 38 bone patients,
5 renal patients and 45 cardiac patients.
Conclusion: This uncertainty in activity exists due to time delay between injection preparation and administration
to patients, as well as due to residual radionuclide that is not injected into patients and remains
in the syringe.
Keywords:
99Mo/99mTcgenerator, radiopharmaceutical, 99mTc, dose calibrator, lead canisters, thyroid patient, cardiac patients,
bone scintigraphy, MDP.
Graphical Abstract
[5]
Chandra, R. Nuclear Medicine Physics :The Basics. 7th ed. Medical Physics, 2012, 39(10), 6525.
[6]
Deutsch, E.A.; Libson, K.; Jurisson, S. Technetium chemistry and technetium radiopharmaceuticals, Progress in Inorganic Chemistry; Lippard, S.J., Ed.; John Wiley & Sons: New York, 1983, pp. 75-139.
[9]
Memon, S.A.; Laghari, N.A.; Qureshi, S.T.; Mehdi, F. Assessment of molybdenum breakthrough levels in molybdenum-99/technetium-99m generators: One year experience at NIMRA Jamshoro Pakistan. Int. J. Radiation Res., 2014, 12(4), 343-346.
[10]
W.M, Haynes., CRC Handbook of Chemistry and Physics. 2014.
[13]
Simon, Cherry; James, Sorenson.; Michael, Phelps Physics.in Nuclear Medicine; USA: Elsevier Health Sciences, 2012.
[14]
Des Plaines, IL IMV Medical Information Division.Nuclear medicine Census Market summary report., 2003.
[15]
Des Plaimes, IL IMV Medical Information Division, Nuclear Medicine Cesus Market Summary Report IMV IMV 2001.
[16]
Subramanian, G.; McAfee, JG.; Blair, RJ.; Kallfelz, FA.; Thomas, FD. Tc-99m methylene diphosphonate a superior agent for skeletal imaging, comparison with other technetium complexes. J. ournal of nuclar medicine., 1975, 16(8), 744-755.
[21]
Kevin, J. Donohoe; Manuel L. Brown; Robert F. Carretta; Robert E. Henkin; Robert E. O’Mara; and Henry D; Royal Society of Nuclear Medicine Procedure Guideline for Bone Scintigraphy, 2003, pp. 205-209.
[22]
Helena, R. Balon, Chair; Edward B. Silberstein; Donald A. Meier; N. David Charkes; Henry D. Royal; Salil D. Sarkar; Kevin J.Donohoe Society of Nuclear Medicine Procedure Guideline for Thyroid Uptake Measurement , 2006; pp. 1-4.
[23]
William, H. Strauss; D. Douglas, Miller; Mark D, Wittry; Manuel D, Cerqueira; Ernest V, Garcia; Abdulmassi S, Iskandrian; Heinrich R, Schelbert; Frans J, Wackers; Helena R, Balon; Otto, Lang; and Josef, Machac. In: Society of Nuclear Medicine Procedure Guideline for Myocardial Perfusion Imaging; , 2008; pp. 155-161.